These show what proportion of light is reflected from the image at each light wavelength. 400nm corresponds to blue light, 700nm corresponds to red light.
Note that most papers appear to reflect a disproportionate amount of blue light. This makes them seem whiter. Some of the measured blue light is probably a result of fluorescence.
These measurements were made from various grey coloured patches printed either from dedicated greyscale targets or extracted from the Gretag Macbeth TC 9.18 RGB target set.
A perfect paper/ink combination would produce a smooth curve between (R=G=B)=0 , L=0 and
(R=G=B)=255, L=100. The shape of the curve would be determined by the "gamma" chosen for the reproduction.
A straight line would be a good start!
The "L" value comes from the "L a b" representation of colour, where "L" corresponds to lightness. L=0 is black, L=100 is white. The other two axes, "a" and "b" represent magenta/green and blue/yellow respectively.
The graph above shows reasonably well behaved results above "L" values of 35. Below this point the graphs become very non-linear - which the colour management system aims to correct. (See below)
The matte papers are not capable of producing the darkest blacks (as Ilford Galerie does). This is partly a result of the fact that the matte surface scatters the light in all directions so the instrument measures the average light reflected whreas a glossy paper produces a simple "specular" reflection which the instrument ignores - in the same way as a real person viewing a glossy print would simply move their viewing angle to eliminate glare from the lights if present.
So the rather poor dark black rendition of the matte papers is not a defect in the paper but an invitable consequence of the matte surface.
Matte papers were printed using the Epson Matte Black ink cartridge.
The Hahnemuehle result gives a larger gamut space then the profile downloaded from their website
(HFAr2400WillTurnerPK.icc dated 18-6-2005).
The calibrated result (for Ilford Galerie) shows improvements in the lowest value of L (the darkest blacks) when using Epson ColorBase calibration.
The calibration also linearises the overall response.